FOGPhotonics,inc-One National Navigation Lab
Fiber optic Gyro coil testing
sevice
REPORT OF TEST
CUSTOMER INFO |
|
Name:U&U
Engineering INC |
|
Add/Zip:uu |
|
DUT |
|
Name:PM fiber
Optic Gyro Coil |
Part
Number:PM15G-80-U13 |
States:Normal |
Serial
No.:0224E01610002 |
Manufacturer:FOGPhotonics,inc |
Our center
is a statutory professional measurement technology institutions, authorized
units: National Defense Science and Technology Industry Bureau |
License
number: National Defense -HKAD-1-009 |
The
center has been approved by the HK National Accreditation Board for
conformity assessment, and the Certificate No. NO.HKASL0546 |
The
centre has been accredited by the Accreditation Committee of the national
defense science and technology industry laboratory, and the certificate
number: NO.LD0089 |
The
center has been approved by the military laboratory, accreditation
certificate No. 20150038 |
All measurement standards of the center have been traced to the highest national or national measurement standards |
Main Equipments Used
in the Test |
||||
Name |
PN/SN |
Serial NO. |
Certificate NO. |
Valid
Date to |
Tunable Laser |
81600B |
DE43500945 |
OFP-1603041 |
2017/03/13 |
Power meter |
81532A |
294G00431 |
GXTC2016-0439 |
2017/03/24 |
OTDR |
FTB-1-S1-8G |
718042 |
OFL-1609019 |
2017/09/06 |
Extinction ratio Meter |
ERM-101 |
203210000135 |
PMC-1512002 |
2016/12/25 |
Return loss tester |
81613A |
DE40500361 |
ORL-1607015 |
2017/07/10 |
High and low
temperature test chamber |
RGDJ-250 |
080109 |
WT2016121 |
2017/06/29 |
Test References: JJG (Electronic) 30903-2008
optical fiber integrated parameter tester verification procedures; JJF
(military) 63-2014 polarization maintaining optical fiber parameter tester
calibration specification |
Test Place: optical
parametric Laboratory
Relative humidity
(R.H.) 26% (Temp) temperature :21.3℃
Unless otherwise stated, the test report shall be dealt with as follows:
This test does not use
sampling or other sampling procedures.
Test items do not have
special use requirements.
When the result of the
test report is wrong, the center will notify the customer in writing.
Test Results:
1.Appearance and
functional examination: normal
2.Polarization
crosstalk distribution:
Polarization crosstalk distribution of fiber coil at temperature of 25℃:
Polarization crosstalk distribution of fiber coil at temperature of -20℃:
Polarization crosstalk
distribution of fiber coil at temperature of 70℃:
3.Insertion Loss:
Tem IL |
1 |
2 |
3 |
Average |
-20℃ |
0.353dB |
0.341dB |
0.352dB |
0.35dB |
25℃ |
0.357dB |
0.345dB |
0.351dB |
0.35dB |
70℃ |
0.352dB |
0.362dB |
0.356dB |
0.36dB |
Extended uncertainty:U=0.07dB (k=2)
4.Return Loss:
Tem RL |
1 |
2 |
3 |
Average |
-20℃ |
49.6dB |
49.8dB |
49.8dB |
49.7dB |
25℃ |
50.2dB |
50.2dB |
50.3dB |
50.2dB |
70℃ |
49.1dB |
49.0dB |
49.0dB |
49.0dB |
Extended uncertainty:U=0.7dB (k=2)
5.Extinction ratio:
Tem PER |
1 |
2 |
3 |
Average |
-20℃ |
17.32dB |
17.33dB |
17.32dB |
17.3dB |
25℃ |
18.04dB |
18.03dB |
17.99dB |
18.0dB |
70℃ |
17.08dB |
17.07dB |
17.07dB |
17.1dB |
Extended uncertainty:U=0.4dB (k=2)
© FOGphotonics, inc. | Company | Products | Technoloy | Contact us